PARAMETERS TECHNICAL

TYPE

MIPS03

MIPS05

MIPS06

MIPS08

MIPS010

Power(kVA)

3.15

5

6.3

8

10

The nominal power of the transformer should be maximum

10 kVA

It should be Single Phase. In case of it being 3 Phase then the L-L Voltage should be 250

Voltage should be 250

PARAMETERS ELECTRIC

 INPUT

phase Configuration

2 in 1 phase + N + PE

nominal Voltage

230V AC

nominal frequency

50/60 Hz

frequency

45-65 (?10%)

OUTPUT

nominal Voltage

230V AC

nominal frequency

50/60 Hz

overload

> 60 s at 150% load

Efficiency in normal operation, the nominal load, linear

>98 %

power factor

0.8

 PARAMETERS TRANSFORMER

The Initial Current value must be

less then 8In 

phase Configuration

2 in 1 phase + N + PE

Temperature measurement of transformer windings

continuous

No Load Condition Current must comply the

Io < % 3 condition. 

 The Short Circuit Voltage must comply the

Uk < % 3 condition

Rate of heat loss in the case of no-load Of nominal capacity

< % 3

The leakage current of the transformer secondary side

<150 ?A

 PARAMETERS STATIC SWITCH

phase Configuration

1Ph+N+PE

voltage control

The two input lines and one output line (for switching)

The scope of the phase voltage changes

220/230/240V AC + -10%

thyristor control

so

Jobs Auto / Manual

so

The range of changes in frequency bypass

47 Hz - 63 Hz

nominal current

50/100 A

Transfer time [ms]

5ms

ENVIRONMENT 

Operating Temperature

-25 ° C ... + 40° C

Humidity

up to 90% (non-condensing)

 

 

STANDARDS

  Standards

IEC60364-7-710:2002-11, section

 

VDE 0107,IEC 61558-2-15, IEC 742

Standard protection class

IP 20

maximum temperature

According to the category B of the transformer insulation system allows a maximum temperature 75°C

 COMMUNICATION

relay contacts

port magistrali CAN

Converting to TCP (optional)